Technical data

Chapter 12 359
GSM and EDGE Analyzer
Optional GSM and EDGE Tx Analyzer Measurement Settings
GSM and EDGE Analyzer
NOTE The disabled gray keys showing your High Limit and your Low Limit in Hz are for
information only. They display the calculated frequency error limits in Hz. The
displayed values (in Hertz) are calculated from your current frequency and your
specified limits in ppm.
Setting TS Power Limits
You can set upper and lower Timeslot Power Limits for your TS Power
Measurement results.
If TS Power Limits are turned On, and your measured TS power is within the limits
you have specified, the numeric results reported in the metrics portion of the screen
will be displayed in green, and the Limit Pass will be indicated by “(P)” after the
2 [Freq Error Limits]
3 [Limits]
4 [On] or [Off]
5 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [ppm]
6 Set the lower limit.
a. [Low Limit]
b. Select Auto or
Manual
c. If you have selected
Manual in Step 6b,
enter the number on
the numeric keypad.
d. If you have selected
Manual in Step 6b,
press [ppm].
If you select Auto, the Low Limit is automatically
set to negative (High Limit). For example, if High
Limit is set to 0.05 ppm, then the Auto Low Limit
is set to –0.05 ppm.
If you select Manual, you must specify the limit in
ppm.
If you switch from Manual to Auto, and then back
to Manual, the previous Manual setting is not
restored.
Step Notes