Technical data

Chapter 13 383
Power Meter
Making an Average Power Measurement
Power Meter
Using the Power Meter Loss Field
Typical base stations exceed the maximum input to most power sensor heads.
Therefore an attenuator (typically a 40 dB 100 W attenuator) is required. The
measured value of the attenuator and connecting cables is entered into this field.
This value is then used when this field is set to
On. When this field is set to On, the
test set will add this loss value to the true measured value indicating the displayed
measured value.
Example: The base station transmits 40 dBm (10 W), the attenuator and RF cabling
15 If you have previously
stored PM loss from an
insertion loss
measurement, go to the
next step.
Or
Press [PM Loss] and
enter the PM loss value.
For information on the effects of the PM Loss
function on the measurement, refer to “Using the
Power Meter Loss field, below.
16 Connect to the base
station.
Connect the power sensor to the signal to be
measured. The connection is typically after the
output of the power amplifier or duplexer. The
connection requires, in addition to the power sensor
and attenuator, a 2 ft. cable and two Type-N barrel
connectors.
Step Notes