Technical data

Chapter 13 385
Power Meter
Making an Average Power Measurement
Power Meter
Setting the Power Meter’s High and Low End-Points
Changing the high and low end points changes the scale resolution, making it easier
to view small changes in power. This does not change the range of the sensor.
Step Notes
1 [Mode]
2[Power Meter]
3 [Setup]
4 On [Resolution] select
1, 2, 3, or 4.
Each time you press this softkey, the selected
option changes.
[1] = 1
[2] = 0.1
[3] = 0.01
[4] = 0.001
Step Notes
1 [Mode]
2[Power Meter]
3 [Setup]
4 [Low End]
5 Enter the minimum value you want
using the numeric keypad.
6[dBm]
7 [Hi End]
8 Enter the maximum scale value
you want using the numeric
keypad.