Technical data

420 Chapter 16
Spectrum Analyzer
Using the Spectrum Analyzer
Spectrum Analyzer
NOTE The Sensitivity setting is coupled to the Noise Correction setting. Changing the
Sensitivity setting from
Low to High will cause the Noise Correction setting to
change from
On to Off. Conversely, if the Noise Correction setting is changed from
Off to On, the Sensitivity setting is changed from High to Low.
Setting the Analyzer Detector
You can set a detector mode for spectrum measurements.
Positive peak detection displays the highest sample level measured during each
sampling period.
Sample detection displays the peak of CW-like signals and maximums and
minimums of noise-like signals.
Negative peak detection displays the lowest sample level measured during each
sampling period.
Average detection displays the average power in each frequency measured
across the spectrum.
2. [Sensitivity]
3.
Low or High
Step Notes
Step Notes
1. [Setup]
2. [Detector]
3. [Peak], [Sample], [Negative Peak],
or [Average]