Using the Spectrum Analyzer
Spectrum Analyzer
14
On [2
nd
Segment] select
On or Off.
Enables you to toggle the measurement limit line
segment function
On (active) and Off (inactive).
• If set to
On, a left and right upper limit line
segment with a pass or fail icon above each
segment is displayed on the screen. This
enables you to set the second segment left and
right limit values. Any measured value that
exceeds the upper limit will result in a failure
indication.
• If set to
Off, neither limit lines nor pass fail
indicators are displayed.
15 Set the second segment
left and right limit
values.
a. On [2
nd
Segment]
select
Left.
b. [Level]
c. Enter the number on
the numeric keypad.
d. [dB]
e. On [2
nd
Segment]
select
Right.
f. Repeat steps b, c, and
d.
Maximum Upper limit: 50 dBm to –150 dBm. If a
Lower Limit is set, the Upper Limit should be set
higher than the Lower Limit)