Technical data
454 Chapter 16
Spectrum Analyzer
Using the Spectrum Emission Mask Measurement
Spectrum Analyzer
Performing a Basic Spectrum Emissions Mask Measurement
Step Notes
1 Select Spectrum Analyzer.
a.
Mode
b. [Spectrum Analyzer/Tools]
c. [Spectrum Analyzer]
2 Enter the channel standard.
a. [Chan Std]
b. Select the standard using the up and
down arrow buttons.
c. [Select]
d. On [Units] select
Chan.
Each time you press the [Units]
menu key, the selected option
changes.
3 Enter the channel number.
a. [Channel]
b. Enter the channel number using the
numeric keypad.
c. Select [Fwd] or [Rev].
4 Select the SEM measurement.
a. [Measurements]
b. [SEM]