Technical data

Chapter 2 47
One Port Insertion Loss
Measuring One Port Insertion Loss
One Port Insertion Loss
Return Loss measurement, and Distance to Fault measurements using a frequency
range that you will set manually, you can perform one calibration for all three
measurements as long as you calibrate over a frequency range that incorporates all
three of your measurements, and as long as your cables do not change. For this
reason, if you are doing Distance to Fault measurements (using a frequency range
that you have set manually) as well as any type of Insertion Loss measurement,
Agilent recommends that you select your cable type before performing calibrations.
Press [Mode], [Antenna/Cable], [Distance to Fault], [Cable Type] to set the cable
type.
The calibration remains valid until you do any one of the following:
set the Distance to Fault frequency range to
Auto. Note that the calibration will
become valid again as soon as you switch from
Auto back to Manual Frequency
Range.
power off the test set
change the start frequency to a new value that lies below the start frequency of
your previous calibration
change the stop frequency to a new value that lies above the stop frequency of
your previous calibration
change any of the jumper leads that you used for the calibration
change any of the attenuators that you used for the calibration
change the type of cable specified under the [Cable Type] menu key.
Performing (and Calibrating) a Basic One Port Insertion Loss
measurement
Select the desired frequency range and calibrate the test set for a One Port Insertion
Loss measurement. To calibrate your test set, you will need the following pieces of
equipment:
Cable, N (male) to N (male), part number 8120-8862 or 8121-1315
Calibrated 50 ohm Open connector, part number 1250-3421
Calibrated 50 ohm Short connector, part number 1250-3424
Calibrated 50 ohm Load connector, part number 1250-3423
NOTE All four of these required accessories are supplied as standard with your Agilent
E7495A/B test set.