Technical data

488 Chapter 18
T1 Analyzer
Making T1 Measurements
T1 Analyzer
Making T1 Measurements
The Agilent E7495A/B provide a simple solution for all your T1 testing needs, from
basic transmission testing to BERT testing. For more information about T1 testing,
refer to “Measurement Results” on page 518.
Performing a Basic T1 Analyzer Measurement
Setup Button Selections
Rx Input: The Primary and Secondary inputs are both affected by this setting.
Terminate - Sets the input impedance to 100 Ohms. Can be performed only
when the T1 circuit is interrupted (connected into a jack that interrupts the
circuit).
Monitor - Sets the input impedance to 100 Ohms, inserts 20 dB gain, and
should be selected when connected to a 'Monitor' jack.
Bridge - Sets the input impedance >1000 Ohms and should be selected when
bridging on to a T1 circuit.
Step Notes
1. [Mode]
2. [Backhaul]
3. [T1 Analyzer]
4. [Get Started/ Test Mode] The tests are listed on the buttons on the right.
Pressing a test mode button will provide you
with information about the test and the test
options available.
5. [Setup] Configure the T1 analyzer for the measurement.
6. [Display] Setup your view of the measurement results.
7. [Control] Modify the measurement setup and inject errors
as you perform the measurement.