Technical data
506 Chapter 18
T1 Analyzer
Making T1 Measurements
T1 Analyzer
13 Select a pattern.
a. [Pattern]
b. Use the up and
down buttons to
select.
c. QRSS
d. [Select]
Set to the pattern sent by the far end.
14 Setup Tx clock.
a. [Tx Clock]
b. [Internal]
This selection is not applicable in Monitor mode
because the Primary Tx signal is a buffered replica
of the signal applied to the Primary Rx jack and is
not affected by this setting. This is the default
selection after test set preset.
15 Setup the Tx LBO.
a. [Tx LBO]
b. [0 dB]
This is a typical level found at the customer
demarcation point.
16 Setup the slip
reference.
a. [Slip Ref]
b. [None]
Disables the frame slip measurement.
17 Setup the second Tx.
a. [Second Tx]
b. [AIS]
Use the default condition.
18 [Display] Enables the [Display] options on the buttons on the
right.
19 Setup the display.
a. [Display]
b. [Status]
c. [Pri]
Enables the [Display] options on the buttons on the
right. Activates only the Primary Status pane. The
inactive channel should be deactivated to stop the
display of invalid measurements.
Step Notes