Technical data
Chapter 2 53
One Port Insertion Loss
Measuring One Port Insertion Loss
One Port Insertion Loss
To set averaging:
NOTE These two steps can be performed in any order. However, if you turn averaging on
and a large number of averages has previously been set, there may be a delay before
you can change the number of averages.
Setting Restart
When you have averaging turned
On and you are in Continuous Sweep mode, you
can restart the averaging by pressing the Restart menu key. The Restart key has no
effect if you are in
Single sweep mode.
To restart averaging:
Step Notes
1 Set the number of averages.
a. [Average]
b. Enter the number of averages
using the numeric keypad.
c. [Enter]
2 Select the type of averaging you
want to apply.
a. [Average/Sweep]
b. [Averaging]
c. [Off], [Running Average],
[Max Hold], [Group
Average], or [Group Max
Average]
Step Notes
1 [Average/Sweep]