Technical data
640 Chapter 21
Safety and Regulatory Information and Specifications
E7495A/B Base Station Test Set Specifications
Safety and Regulatory Information
and Specifications
Instrumentation Accuracy:
Absolute:
Relative:
± 0.02 dB (log) or ± 0.5% (linear) Add the
corresponding power sensor linearity percentage.
± 0.04 dB (log) or ± 1.0% (linear) Add the
corresponding power sensor linearity percentage.
Zero set accuracy: Zero set is the digital zero with an N8482A sensor
±63nW
Power reference accuracy:
Power output:
Accuracy:
VSWR:
1.00 mW (0.0 dBm) traceable to the U.S. National
Institute of Standards and Technology (NIST).
± 1.2% worst case (± 0.9% rss) for one year
< 1.08
External attenuator:
Max power:
Attenuation:
100 watts
40 dB ± 0.5 dB
Power meter Option 600 with Agilent N8482A-CFT power sensor
(also supports Agilent 8481A/8482A, N8481A-CFT and 8481D power sensors)
Frequency Range: 100 kHz to 6 GHz
VSWR
100 kHz to 300 kHz:
300 kHz to 1 MHz:
1 MHz to 2 GHz:
2 GHz to 6 GHz:
< 1.52
< 1.12
< 1.06
< 1.07
Power linearity: - 1 dBm to +15 dBm; ± 0.52%
+15 dBm to +20 dBm; ± 0.80%
Maximum power: 300 mW average. 15W Peak for 2 μs
Measurement noise: < 114 nW
Zero drift: < ± 7 nW
Table 2 E7495A/B Base Station Test Set Option Specifications