Technical data

Chapter 3 95
1xEV-DO Analyzer and Over Air Test
Optional 1xEV-DO Tx and Over Air Analyzer Measurement Settings
1xEV-DO Analyzer and Over Air
Test
To Set the Measurement Frequency Error Limits:
Step Notes
1 [Setup]
2 [Limits]
3 On [Freq Error Limits]
select
On or Off.
Enables you to toggle the measurement limits
function On (active) and Off (inactive).
If set to
On, the Frequency Error will change
from black to red (fail), or to green (pass), and
will indicate either “P” (pass) or “F” (fail). If
the measured value exceeds either the upper
limit or the lower limit, the words ‘ALL PASS’
in the bar at the bottom of the screen will
change to ‘FAILURE’.
•If set to
Off, the channel freq/number value will
be displayed in blue. The bar at the bottom of
the screen will indicate measurement in
progress.
4 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [GHz], [MHz],
[kHz], or [Hz]
5 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [GHz], [MHz],
[kHz], or [Hz]