Technical data
Chapter 3 97
1xEV-DO Analyzer and Over Air Test
Optional 1xEV-DO Tx and Over Air Analyzer Measurement Settings
1xEV-DO Analyzer and Over Air
Test
Setting the Idle On/Off Ratio Limits
Step Notes
1 Press [On/Off Ratio
Limits].
2 Press [Limits]. Enables you to toggle the Idle On/Off Ratio Limits
function
On (active) and Off (inactive).
• If set to
On, the displayed On/Off Ratio metric
will change from blue to red (fail), or to green
(pass), and will indicate either “P” (pass) or “F”
(fail). If the measured value exceeds the lower
limit, the words ‘ALL PASS’ in the bar at the
bottom of the screen will change to
‘FAILURE’.
•If set to
Off, the Idle Power and Non Idle Power
values will be displayed in blue. The bar at the
bottom of the screen will indicate measurement
in progress.
3 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [Enter]
Lower Limit range: –200 to 200 dB