Technical data
Chapter 3 99
1xEV-DO Analyzer and Over Air Test
Optional 1xEV-DO Tx and Over Air Analyzer Measurement Settings
1xEV-DO Analyzer and Over Air
Test
Setting the Estimated MAC Rho Lower Limit
Step Notes
1 Press [Est MAC Rho
Limit].
2 Press [Limits]. Enables you to toggle the Estimated MAC Rho
measurement’s lower limit function
On (active) and
Off (inactive).
• If set to
On, the displayed Estimated MAC Rho
metrics will change from blue to red (fail), or to
green (pass), and will indicate either “P” (pass)
or “F” (fail). If the measured value falls below
the lower limit, the words ‘ALL PASS’ in the
bar at the bottom of the screen will change to
‘FAILURE’.
•If set to
Off, the Estimated MAC Rho value will
be displayed in blue. The bar at the bottom of
the screen will indicate measurement in
progress.
3 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [Enter]
Lower Limit range: 0 to 1