Specifications
26
The PNA-L is an integrated vector network analyzer equipped with 
a built-in S-parameter test set, one or two synthesized sources 
used for device stimulus, a hard disk drive, USB interfaces, and an 
8.4” LCD color touch screen display. 40 and 50 GHz models have 
2.4 mm ruggedized male 50-ohm connectors, while all other mod-
els have 3.5 mm ruggedized male 50-ohm connectors. Included 
with each instrument is a mouse, keyboard (U.S.) and a 1-year 
return-to-Agilent service warranty.
Test set and power configuration options
•  Standard test set and power range (Option xx0)
  The 13.5 and 20 GHz versions are available in 2- or 4-ports.
•  Configurable test set and extended power range (Option xx5)
  Adds front panel access loops and one or two 60 dB step  
  attenuators as shown in the figures below. This provides the 
  capability to improve instrument sensitivity for measuring 
 low-level signals, to reverse the directional couplers to achieve 
  even more dynamic range or to add components and other  
  peripheral instruments for a variety of measurement applications. 
 The 13.5 and 20 GHz versions are available in 2- or 4-ports.
•  Configurable test set, extended power range and internal 
  second source (Option x46)
  Available with 4-port models only, this option adds an internal 
  second source, nine front panel access loops and two 60 dB 
  step attenuators as shown in the figure below. This provides an 
  additional signal (fixed or swept) for two-tone third-order-intercept 
  (TOI) and intermodulation testing of amplifiers; or it can be used 
  as a fast swept-LO signal for fixed-IF testing of mixers and 
  converters. In either case, sweep speed is more than twenty times 
  faster than using an external source (Option 080 recommended. 
  Please note, Option 080 is needed in order to have independent 
  control of the two internal sources.).
2-port standard test set
2-port configurable test set and 
extended power range 
Source
Switch/splitter
Reference 
receiver
Reference 
receiver
Measurement 
receivers
Port 1 Port 2
A B
R1 R2
Source
Switch/splitter
Reference 
receiver
Reference 
receiver
Measurement 
receivers
Port 1 Port 2
60 dB
A
60 dB
B
R1 R2
Source
out
Rcvr
R1
in
SRC
out
Cplr
thru
Cplr
arm
Rcvr
A
in
Rcvr
R2
in
Source
out
Rcvr
B
in
Cplr
arm
Cplr
thru
SRC
out
Source
Switch
PORT 4
Coupler arm
Receiver D in
Source out
Coupler thru
PORT 3
Coupler arm
Receiver C in
Source out
Coupler thru
PORT 2
Coupler thru
Source out
Receiver B in
Coupler arm
PORT 1
Coupler thru
Source out
Receiver A in
Coupler arm
D
R
Reference
receiver
Source out
Receiver R in
C
A
B
60 dB
Source 1
Source 2
Test port 2Test port 1
Test port 4
Test port 3
60 dB 60 dB
R
A B C D
4-port standard test set
4-port configurable test set and 
extended power range 
4-port configurable test set, extended 
power range and internal second source
Source
Switch
PORT 4
PORT 3
PORT 2
PORT 1
R
Reference
receiver
C
A
B
D
PNA-L Series Network Analyzer 