Instruction manual

Chapter 4 101
Troubleshooting
Troubleshooting Using Diagnostic Test
4. Troubleshooting
Test equipment required for diagnostic test
Table 4-2 shows the equipment required for performing the diagnostic test.
To Execute the Diagnostic Test
To isolate faulty board assembly in analog section, execute the diagnostic test in
accordance with the following procedure. The test procedure needs to be performed using a
mouse and an external keyboard in addition to the front panel keys.
NOTE To perform the diagnostic test properly, the following conditions must be met:
1. Environmental temperature: 23C 5C
NOTE Do not operate front panel keys, keyboard and mouse during the diagnostic test. Changing
the instrument settings while the diagnostic test is in progress will cause incorrect test
results.
Step 1. Connect a mouse and an external keyboard to the E5061A/E5062A’s rear panel connector.
Step 2. Press key.
Step 3. Press
Load Project to select Load Project function."Open" dialog box will be displayed as
shown in Figure 4-10.
8 Receiver Compression Test
9 Receiver IF Ranging Test
10 Handler I/O Test
Table 4-1 Diagnostic test group menu
Table 4-2 Required equipment
Required test equipment Qty Recommended model
Coaxial cable with N type (m) connectors,
61 cm (24 in)
1 N6314A, 8120-8898(for 75 )
Calibration kit 1 85032B/F, 85036B(for 75 )
Type-N coaxial Short termination 2 part of 85032B/F, 85036B(for 75
)
Power splitter 1 0955-0751, 0955-0752(for 75 )
Handler I/O test kit 1 E5070-65001