222 Appendix C
Firmware Update
Firmware Update
3. Enter the password e506xa into the Password box.
4. Open dialog box apperas. Confirm that
E5062.msi is displayed at the File name
column. Press the
Open button, and the Windows instoller will appear.
5. After the disk 1 installation is finished, insert the firmware installation disk 2 (2 of 4)
into the FDD of the E5061A/E5062A, then click
OK.
6. After the disk 2 installation is finished, insert the firmware installation disk 3 (3 of 4)
into the FDD of the E5061A/E5062A, then click
OK.
7. After the disk 3 installation is finished, insert the firmware installation disk 4 (4 of 4)
into the FDD of the E5061A/E5062A, then click
OK.
Step 5. After the disk 4 installatio is fininshed, the instrument will be restarted and the
measurement view will appear.
Step 6. Press -
Firmware Revision. Confirm the firmware revision you have just installed.