Instruction manual

Chapter 2 45
Performance Tests
E5061A Performance Test Record (for Option 150/250)
2. Performance Test
Uncorrected System Performance Test
@ Correction: Off
S21 or S11
S12 or S22 (Option 250 only)
-75 0.270 0.028
-80 0.371 0.028
-85 0.546 0.030
-90 0.841 0.030
-95 1.334 0.032
-100 2.141 0.032
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
System
performance
Frequency range [Hz] Test limit [dB] Test result [dB]
Directivity 300 k - 100 M -25
Directivity 100 M - 1.5 G -25
Source match 300 k - 100 M -25
Source match 100 M - 1.5 G -25
Load match 300 k - 100 M -15
Load match 100 M - 1.5 G -15
Reflection Tracking 300 k - 100 M 1.0
Reflection Tracking 100 M - 1.5 G 1.0
Transmission
Tracking
300 k - 100 M 1.0
Transmission
Tracking
100 M - 1.5 G 1.0
System
performance
Frequency range [Hz] Test limit [dB] Test result [dB]
Directivity 300 k - 100 M -25
Directivity 100 M - 1.5 G -25
Source match 300 k - 100 M -25
Source match 100 M - 1.5 G -25