Instruction manual

Chapter 2 47
Performance Tests
E5061A Performance Test Record (for Option 175/275)
2. Performance Test
E5061A Performance Test Record (for Option 175/275)
Agilent Technologies E5061A RF Network Analyzer
Frequency Accuracy Test
RF Output Level Accuracy and Flatness Test
Level Accuracy Test (@ Port 1)
Level Flatness Test (@ Port 1) (at 0dBm, relative to 50 MHz reference)
Serial Number: Option:
Temperature: C Test Date:
Humidity: % R.H. Tested by:
Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement
uncertainty [Hz]
300 k 1.5 0.056
50 M 250 9
1.5 G 7.5 k k 0.28 k
Power level
[dBm]
Frequency
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
0 50 M 0.8 0.10
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
300 k 1.0 0.34
1.5 M 1.0 0.10
3.00 M 1.0 0.10
7.00 M 1.0 0.10
15.0 M 1.0 0.10
275 M 1.0 0.10
575 M 1.0 0.10
875 M 1.0 0.10
1.175 G 1.0 0.10