Instruction manual

58 Chapter 2
Performance Tests
E5062A Performance Test Record (for Option 150/250)
Level Accuracy Test (@ Port 2) (Option 250 only)
Level Flatness Test (@ Port 2) (at 0dBm, relative to 50 MHz reference) (Option 250
only)
1.175 G 1.0 0.07
1.5 G 1.0 0.08
1.775 G 1.0 0.08
2 G 1.0 0.16
2.375 G 1.0 0.16
2.675 G 1.0 0.16
3 G 1.0 0.16
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level
[dBm]
Frequency
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
0 50 M 0.8 0.07
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
300 k 1.0 0.27
1.5 M 1.0 0.07
3.00 M 1.0 0.07
7.00 M 1.0 0.07
15.0 M 1.0 0.07
275 M 1.0 0.07
575 M 1.0 0.07
875 M 1.0 0.07
1.175 G 1.0 0.07
1.5 G 1.0 0.08
1.775 G 1.0 0.08
2 G 1.0 0.16
2.375 G 1.0 0.16
2.675 G 1.0 0.16