Instruction manual

Chapter 2 65
Performance Tests
E5062A Performance Test Record (for Option 150/250)
2. Performance Test
Dynamic Accuracy Test
@ Reference power level: -10 dBm
Direction: S21
Direction: S12 (Option 250 only)
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 0.303 0.011
5 0.144 0.011
0 0.090 0.011
-5 0.068 0.011
-15 0.065 0.011
-20 0.073 0.011
-25 0.080 0.015
-30 0.087 0.015
-35 0.095 0.019
-40 0.103 0.019
-45 0.111 0.022
-50 0.121 0.022
-55 0.134 0.024
-60 0.150 0.024
-65 0.174 0.026
-70 0.211 0.026
-75 0.270 0.028
-80 0.371 0.028
-85 0.546 0.030
-90
0.841 0.030
-95 1.334 0.032
-100 2.141 0.032
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 0.303 0.011
5 0.144 0.011