Troubleshooting
How to exit from the E5061A/E5062A Measurement View
4. Troubleshooting
How to exit from the E5061A/E5062A Measurement View
You need to exit from the E5061A/E5062A Measurement View to perform some
troubleshooting. The following is the procedure to exit from the E5061A/E5062A
Measurement View.
Step 1. Connect the mouse and external keyboard to the connectors on the E5061A/E5062A rear
panel.
Step 2. Turn the instrument on.
Step 3. Press key.
Step 4. Click
Service Menu - Service Function. Password dialog box opens as shown in Figure
4-1.
Figure 4-1 Password dialog box
Step 5. Enter the password
kid in the Password box and then click "OK" button.
Step 6. Click
Exit in Service Functions Menu. Then the E5061A/E5062A exit the Measurement
View, then.windows desktop screen appears with 4 icons (My Computer, Recycle Bin, My
Network Places, and Network Analyzer.)
NOTE If you wish to return to the Measurement View, double-click "Network Analyzer" icon.
NOTE If you need to shut down the E5061A/E5062A and again turn on, perform in accordance
with the following procedure.
a. To get "Start" menu bar displayed, move the pointer to the bottom of the screen with
mouse.
b. Click "Start" and "Shut Down..." in the pull down menu. "Shut Down Windows" dialog
box opens.
c. Select "Shut down" button in the pull down menu.
d. Click "OK" button in the dialog box.