Specifications

ESG Family Signal Generators Performance Tests
22. Dual Arbitrary Waveform Generator Check (Option UND only)
Calibration Guide 4-43
3. Verify that Channel 1 and Channel 2 are both displaying triangular waveforms
approximately 180 degrees out of phase and that they do not have any discontinuities.
Refer to the illustration below.
Frequency is approximately 64 kHz.
Amplitude is approximately 1.9 V
p-p
.
Both waveforms are symmetrical around 0 volts (the average voltage is
approximately 0V).