Specifications
ESG Family Signal Generators Performance Tests
22. Dual Arbitrary Waveform Generator Check (Option UND only)
Calibration Guide 4-43
3. Verify that Channel 1 and Channel 2 are both displaying triangular waveforms
approximately 180 degrees out of phase and that they do not have any discontinuities.
Refer to the illustration below.
• Frequency is approximately 64 kHz.
• Amplitude is approximately 1.9 V
p-p
.
• Both waveforms are symmetrical around 0 volts (the average voltage is
approximately 0V).