Specifications

Adjustments ESG Family Signal Generators
Adjustment Relationships
5-2 Calibration Guide
Adjustment Relationships
Anytime an adjustment is made to the signal generator other related adjustments may be
affected. For optimal performance, whenever an adjustment is performed, the related
adjustments should also be performed.
Analog Bus Related Adjustments
The following adjustments are optimized if the Analog Bus ADC Calibration has been run:
VCO Bias Potentiometer Adjustment
FM Inverting Amplifier Offset Calibration
ALC Modulation Drive Bias Calibration
VBLO Mixer Bias Calibration
Level Meter Calibration
I/Q Gain/Offset/Quadrature Calibration (ESG-D only)
I/Q Impairment Adjustment (ESG-D only)
Synthesizer Potentiometer Related Adjustments
The Lock Angle potentiometer is preset fully counter-clockwise in the VCO Bias
Adjustment. Both adjustments should always be performed together.
VCO Bias Potentiometer Adjustment
Lock Angle Adjustment
AM Accuracy Related Adjustments
AM Audio Path Offset Calibration
AM Gain Calibration
I/Q Modulation Accuracy Related Adjustments
VBLO Mixer Bias Calibration
I/Q Gain/Offset/Quadrature Calibration (ESG-D only)
I/Q Impairment Adjustment (ESG-D only)
Baseband Generator I/Q Gain and Offset Adjustment (Options UN3 or UN4 only)
Burst Modulation Accuracy Related Adjustments
Burst Modulator Calibration (ESG-D only)
Burst Audio Path Gain Calibration (ESG-D only)