Specifications

Performance Tests ESG Family Signal Generators
9. Harmonic, Subharmonic, and Nonharmonic Spurious Signals
4-20 Calibration Guide
9. Harmonic, Subharmonic, and
Nonharmonic Spurious Signals
These automated tests verify that the harmonic, subharmonic, and nonharmonic spurious
signals are within specifications. The signal generator’s output signal is set to values
where harmonic and spurious signal performance problems are most likely to occur. Then
the spectrum analyzer checks the level of the various harmonic and spurious signals and
compares each measurement to a signal of known frequency and amplitude.
Recommended Equipment
HP/Agilent 8563E Spectrum Analyzer
HP/Agilent 8491A/B Attenuator (10 dB) Option 010
Equipment Setup
Figure 4-18 Harmonic, Subharmonic, and Nonharmonic Spurious Signals Tests Setup