Basic Signal Generator Operations
Basic Signal Generator Operations
Cannot Turn Off Help Mode
1. Press Utility > Instrument Info/Help Mode
2. Press
Help Mode Single Cont until Single is highlighted.
The signal generator has two help modes; single and continuous.
When you press
Help in single mode (the factory preset condition), help text is provided for the
next key you press. Pressing another key will exit the help mode and activate the key’s
function.
When you press
Help in continuous mode, help text is provided for the next key you press and
that key’s function is also activated (except for
Preset). You will stay in help mode until you
press
Help again or change to single mode.
No RF Output
Check the RF ON/OFF annunciator on the display. If it reads RF OFF, press RF On/Off to toggle
the RF output on.
The Power Supply has Shut Down
If the power supply is not working, it requires repair or replacement. There is no
user-replaceable power supply fuse. Refer to the Service Guide for instructions.
Signal Loss While Working with Mixers
If you experience signal loss at the signal generator’s RF output during low-amplitude coupled
operation with a mixer, you can solve the problem by adding attenuation and increasing the
RF output amplitude of the signal generator.
Figure 16-1 on page 464 shows a hypothetical configuration in which the signal generator
provides a low amplitude signal to a mixer.