Specifications
4-18 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Operation ESG Family Signal Generators
Understanding the Reference Measurement Channel Option 200
Understanding the Reference Measurement Channel
The signal generator provides one-button setup capabilities for transport channel
configuration. The dedicated physical channel DCH (downlink) or DPDCH (uplink) is
predefined by pressing the
Ref Measure Setup softkey (or sending the appropriate SCPI
commands). For downlink, at least one physical channel needs to be set to DPCH to
activate the
Ref Measure Setup softkey. Also, to activate the Config Transport softkey, one of
the reference measurement rates must be selected or the DPCH Data field value must be
set to Transport CH. For uplink,
DPDCH and RMC 12.2 kbps are the default selections,
respectively.
Table 4-12 and Table 4-13 describe the downlink and uplink reference measurement
channel (RMC) configurations generated by pressing the
Ref Measure Setup softkey after
signal generator preset. Transport channel parameters can be modified in a table editor by
pressing the
Config Transport softkey, then moving the cursor to the desired data field and
pressing
Edit Item. DPCH and DPDCH parameters can be individually modified in a table
editor by pressing
PhyCH Setup, then moving the cursor to the desired data field and
pressing
Edit Item.
Table 4-12 The Downlink RMC Predefined DPCH Configuration
Parameter DPCH Values at Specified Reference Measurement Channel
12.2 kbps 64 kbps 144 kbps 384 kbps UDI 64 AMR 12.2 UDI
ISDN
Power 0.00 dB 0.00 dB 0.00 dB 0.00 dB 0.00 dB 0.00 dB 0.00 dB
Channel
Code
a
a. ESG Channel # 1 default shown. Default channel code for Channels 2 through 4 are based on the
equation: Default Channel Code = (Channel #) + 5.
6666666
SecSrc Code
OS
0000000
TPC Pat
Steps
1111111
Data
b
b. If a transport configuration parameter is changed using the Config Transport
softkey and the table editor, the Data field reverts to Transport CH, showing that it no longer con-
tains a specified reference measurement channel.
Ref 12 Ref 64 Ref 144 Ref 384 ISDN64 AMR 12 ISDN
Symbol
Rate
c
c. Symbol rate is not user-selectable. It is coupled to slot format. To change symbol rate, use the
appropriate slot format.
30.00
ksps
120.0
ksps
240.0
ksps
480.0
ksps
120.0
ksps
30.0 ksps 120.0
ksps
TFCI
Pattern
00000000 00000000 00000000 00000000 00000000 00000000 00000000
TPC Pattern Up/Down Up/Down Up/Down Up/Down Up/Down Up/Down Up/Down
Slot Format 11 13 14 15 13 8 13
Time Offset0000000