Specifications

4-20 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Operation ESG Family Signal Generators
Understanding the Single-ESG Downlink Configuration Option 200
Understanding the Single-ESG Downlink Configuration
Capabilities
The Option 200 single-ESG setup provides four fully coded 3GPP channels for use in
testing 3GPP mobile receiver designs.
Channel Types
The single-ESG setup can provide up to four of the following downlink channel types:
primary synchronization channel (PSCH)
secondary synchronization channel (SSCH)
primary common control physical channel (PCCPCH)
common pilot channel (CPICH)
dedicated physical channel (DPCH)
paging indicator channel (PICH)
orthogonal channel noise simulator (OCNS)
A single ESG can provide a control configuration (PSCH, SSCH, PCCPCH, and CPICH) or
an information data transmission configuration (DPCH and PICH, with optional OCNS).
For valid channel type/channel number, see “Valid Channel-Type-to-Channel-Number
Assignments” on page 4-7. For valid channel combinations, see “Valid Channel
Combinations per ESG” on page 4-7.
Reference Measurement Channels
Option 200 provides reference measurement channels at 12.2, 64, 144, and 384 kbits per
second. This option also provides transport layer channelization for AMR 12.2 (adaptive
multi-rate) and UDI 64 (unrestricted digital information) protocols.
System Triggering and Synchronization
This configuration utilizes internal I/Q triggering. The rear panel EVENT 2 BNC
connector provides a system synchronization signal in the form of a chip-wide pulse that
resets the system every time a parameter change that requires system re-synchronization
is made. For increased measurement accuracy, the signal generators rear panel 10 MHZ
OUT frequency reference can be utilized by other instruments in the test system.