Specifications
Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide 2-7
ESG Family Signal Generators Using Functions
Option 200 Two ESG Downlink Measurement Setup
Connect the Equipment
1. Refer to Figure 2-2 and connect the two ESGs as shown:
NOTE This denotes a short cable.
This denotes a long cable.
Figure 2-2 ESG Downlink Equipment Setup
2. Turn on the function generator and set the following characteristics:
wave type: square wave
amplitude: 2.5 V
rms
frequency: 3.84 MHz
DC offset: +1.25 V
3. Turn on the signal generators and the transmitter tester.