Specifications
2-76 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Using Functions ESG Family Signal Generators
Uplink Measurement Setup Option 200
Uplink Measurement Setup
This section will guide you through setting up the ESG and measuring the output using
the E4406A VSA Transmitter Tester.
This setup provides two 3GPP channels with real-time signal generation capability, which
are used to test 3GPP base station (BS) receiver designs.
Required Equipment
• one ESG-D Series Signal Generator with Option 200
• one E4406A VSA Series Transmitter Tester with Option BAF
• BNC cables as needed (see the following figure)
Refer to Figure 2-15 and make the necessary connections between the ESG and the
E4406A.
Figure 2-15 ESG Uplink Measurement Setup
Configuring the ESG
NOTE The following signal generator settings are used for example purposes.
Change them as needed for your device under test (DUT).
1. Press
Preset.
2. Press
Frequency > 1.950 > GHz and then Amplitude > –10 > dBm.
3. Press
Mode > Real Time I/Q BaseBand > CDMA > W-CDMA.
4. Press
Link Down Up until Up is highlighted.
5. Press
UE Setup.
This opens the user equipment (UE) setup editor. The highlight bar can be moved by
rotating the front panel knob or pressing the arrow keys.