Specifications
2-34 Programming Guide
Programming Commands and Examples ESG Family Signal Generators
:LIST Subsystem
:LIST Subsystem
The list subsystem is used to set the controls and the parameters associated with
frequency and/or power sweeps.
Dwell List
:LIST:DWELl <val>{, <val>}
:LIST:DWELl?
This command defines a list of dwell times. The list is stored in a file (DWEL_FILE) which
is only reinitialized upon power-up preset. Each value specifies the minimum time that the
sweep engine waits at that point’s frequency and/or power setting before going to the next
setting. Dwell times are only used when the point trigger source is set to Immediate.
Dwell List Points Query
:LIST:DWELl:POINts?
This command queries the signal generator for the number of points in the dwell list.
Dwell List Type
:LIST:DWELl:TYPE LIST|STEP
:LIST:DWELl:TYPE?
This command sets the type of dwell for a list sweep. The choices are List or Step. Use this
command to toggle the dwell time for the list sweep points between the values defined in
the list sweep and the value set for step sweep. Choose
LIST to sweep with dwell times that
you have defined in the list sweep. Choose
STEP to sweep each point in the list with a dwell
time set via the :SWEep:DWELl <val> command.
*RST Value: List
Frequency List
:LIST:FREQuency <val>{, <val>}
:LIST:FREQuency?
This command defines the frequency list. Each value specifies the frequency that the
sweep engine uses when sweeping frequency. This list is stored in a file (FREQ_FILE),
which is only
reinitialized upon power-up preset.
Frequency List Points Query
:LIST:FREQuency:POINts?
This command queries the signal generator for the number of points in the frequency list.