Specifications

Programming Guide 2-53
ESG Family Signal Generators Programming Commands and Examples
:POWer Subsystem
RF Output Automatic Leveling Circuitry (ALC) Search State
:POWer:ALC:SEARch ON|OFF|1|0|ONCE
:POWer:ALC:SEARch?
This command toggles between the auto and manual modes of power search mode. The
choices are On (1) or Off (0). Power search is an internal calibration routine used to achieve
calibrated output power when the ALC is off. When you turn power search on, power
search will execute automatically with each change to the RF frequency or power, and also
with each change to the AM, burst, pulse, or I/Q modulation state. You do not need to
disconnect modulation inputs, however modulation is disabled during power search. If you
are using external I/Q modulation, the power search routine assumes that the I/Q input
level is = 0.5 V
rms
. Actual output amplitude will scale directly with . When
power search is set to Off, the power search routine is not run unless the command
[:SOURce]:POWer:ALC:SEARch ONCE is sent.
*RST Value: Off
RF Output Automatic Leveling Circuitry (ALC) Search Reference
:POWer:ALC:SEARch:REFerence FIXed|MODulated
:POWer:ALC:SEARch:REFerence?
This command lets you set either fixed or modulated modes of power search. Power search
is an internal calibration routine used to achieve calibrated power when the ALC is off.
Fixed power search mode uses a 0.5 volt reference. When you set the power search mode to
modulated, the current I/Q modulation remains active and the reference is the RMS value
of the current I/Q modulation.
*RST Value: Fixed
RF Output Automatic Leveling Circuitry (ALC) State
:POWer:ALC:STATe ON|OFF|1|0
:POWer:ALC:STATe?
This command sets the operating state of the signal generator’s RF output automatic
leveling circuitry. The choices are On (1) or Off (0).
*RST Value: On