:AM Subsystem
Internal Amplitude Modulation Sweep Trigger
:AM[1]|2:INTernal[1]:SWEep:TRIGger IMMediate|BUS|EXTernal|KEY
:AM[1]|2:INTernal:SWEep:TRIGger?
This command selects the trigger for the amplitude modulation sweep.
*RST Value: Immediate
Internal Amplitude Modulation Waveform
:AM[1]|2:INTernal[1]:FUNCtion:SHAPe SINE|TRIangle|SQUare|RAMP|NOISe|
DUALsine|SWEPtsine
:AM[1]|2:INTernal:FUNCtion:SHAPe?
This command allows you to assign an AM waveform to your
AM Path 1 and AM Path 2
configurations. Select from sine, triangle, square, ramp, noise, dual sine, and swept sine
waveforms. Notice that your waveform selection applies only to whichever AM path
configuration you have currently selected.
Using the command :AM[1]|2:INTernal[1]:FUNCtion:SHAPe SINE lets you specify sine as
the amplitude modulation waveform for the
AM Path 1 and AM Path 2 configurations. Notice
that the selected waveform applies only to whichever AM path configuration you have
currently selected.
Using the command :AM[1]|2:INTernal[1]:FUNCtion:SHAPe TRIangle lets you specify
triangle as the amplitude modulation waveform for the
AM Path 1 and AM Path 2
configurations. Notice that the selected waveform applies only to whichever AM path
configuration you have currently selected.
Using the command :AM[1]|2:INTernal[1]:FUNCtion:SHAPe SQUARe lets you specify
square as the amplitude modulation waveform for the
AM Path 1 and AM Path 2
configurations. Notice that the selected waveform applies only to whichever AM path
configuration you have currently selected.
Using the command :AM[1]|2:INTernal[1]:FUNCtion:SHAPe RAMP lets you specify ramp
as the amplitude modulation waveform for the
AM Path 1 and AM Path 2 configurations.
Notice that the selected waveform applies only to whichever AM path configuration you
have currently selected.
Using the command :AM[1]|2:INTernal[1]:FUNCtion:SHAPe NOISe lets you specify noise
as the amplitude modulation waveform for the
AM Path 1 and AM Path 2 configurations.
Notice that the selected waveform applies only to whichever AM path configuration you
have currently selected.
Using the command :AM[1]|2:INTernal[1]:FUNCtion:SHAPe DUALsine allows you to set
the dual sine as the amplitude modulation waveform for the
AM Path 1 and AM Path 2
configurations. In this mode you can set the AM rates for two separate tones. In addition,
you can set a percentage of the tone 2 AM depth to the total AM depth.