Specifications
2-12 Programming Guide
Programming Commands and Examples ESG Family Signal Generators
:CALibration Subsystem
:CALibration Subsystem
The calibration subsystem is used to set the controls and the parameters associated with
instrument calibration.
DCFM/DCΦM Calibration
:CALibration:DCFM
There is no query for this command.
This command initiates a DCFM or DC
ΦM calibration (depending on which kind of
modulation is currently active) and stores the results in the instrument’s firmware. This
calibration eliminates the offset in phase (or frequency, as appropriate) modulation so that
the carrier phase (or frequency) remains the same with no modulation applied. An external
DC-coupled phase (or frequency) modulation must be active when the calibration is
performed. After calibration, the DC signal produces no modulation.
There are no initial values associated with this command.
Default I/Q Calibration
:CALibration:IQ:DEFault
The is no query for this command.
This command will restore the original factory calibration data for the internal I/Q
modulator. The latest calibration data from the user calibration is discarded. Although the
factory calibration data may not be optimized for the signal generator’s present status, the
data can be used as a starting point if the current data is lost (if the I/Q calibration is
unable to execute).
There are no initial values associated with this command.
Full I/Q Calibration
:CALibration:IQ:FULL
There is no query for this command.
This command sets and performs a full-frequency range (regardless of the start and stop
frequency settings) I/Q calibration and stores the results in the instrument’s firmware.
There are no initial values associated with this command.