User`s guide
92 Chapter 3
Basic Digital Operation
Using Waveform Markers
Accessing Marker Utilities
Use the following procedure to display the marker parameters. This procedure uses the Dual ARB 
player, but you can access the marker utilities through the 
ARB Setup softkey in all ARB formats.
1. Select the ARB waveform player:
press 
Mode > Dual ARB
2. Press  ARB Setup > Marker Utilities.
Sample range begins on first point of signal
Negative range set between signal and 
off time
Example of Incorrect Use
Waveform: 1022 points
Marker range: 110-1022
Marker polarity: Negative
This figure shows that a negative polarity marker goes low during 
the marker on points; the marker signal goes high during the off 
points. The ALC samples the waveform during the off marker 
points.
Marker On
Marker On
Sampling both on and off time sets the modulator circuitry 
incorrectly for higher signal levels. Note the increased amplitude 
at the beginning of the pulse.
Marker
Off
Marker On
Marker
Marker
Off
On