User`s guide
Chapter 3 107
Basic Digital Operation
Triggering Waveforms
NOTE In the real- time Custom mode, the behavior is reversed: when the gating signal is high, you 
see the modulated waveform.
Using Segment Advance Triggering
Segment advance triggering enables you to control the segment playback within a waveform sequence. 
The following example uses a waveform sequence that has two segments.
If you have not created and stored a waveform sequence, complete the steps in the sections, 
“Creating Waveform Segments” on page 84, and “Building and Storing a Waveform Sequence” on 
page 85.
1. Preset the signal generator.
2. Configure the RF output:
• Set the desired frequency.
• Set the desired amplitude.
• Turn on the RF output.
3. Select a waveform sequence for playback:
a. Press 
Mode > Dual ARB > Select Waveform.
b. Highlight a waveform sequence file.
c. Press 
Select Waveform.
4. Select the waveform trigger method and trigger source:
a. Press  Trigger >  Segment Advance.
b. Press 
Trigger > Trigge r  Setup  and note that the Seg Advance Mode softkey displays the default 
selection (Continuous), which is the selection used in this example.
c. Press 
Trigger Source >  Trigger Key.
Externally Applied Gating Signal
Gate Active = High
RF OutputModulating Waveform