User`s guide
Chapter 4 121
Optimizing Performance
Using External Leveling
Configure the Signal Generator
1. Press 
Preset.
2. Press 
Frequency > 10 > GHz.
3. Press 
Amplitude > 0 > dBm.
4. Press 
RF On/Off.
5. Press 
Leveling Mode > Ext Detector.
This deactivates the internal ALC detector and switches the ALC input path to the front panel 
ALC INPUT connector. The EXT indicator is activated in the AMPLITUDE area of the display.
NOTE For signal generators with Option 1E1, notice that the ATTN HOLD (attenuator hold) 
annunciator is displayed. During external leveling, the signal generator automatically 
uncouples the attenuator from the ALC system for all external leveling points. While in 
this mode, the RF output amplitude adjustment is limited to −20 to +25 dBm, the 
adjustment range of the ALC circuitry. For more information, see “External Leveling with 
Option 1E1 Signal Generators” on page 122.
6. Observe the coupling factor printed on the directional coupler at the detector port. Typically, this 
value is −10 to −20 dB.
Enter the positive dB value of this coupling factor into the signal generator.
7. Press 
More (1 of 2) > Ext Detector Coupling Factor > 16 (or the positive representation of the value listed at 
the detector port of the directional coupler) > 
dB.
Leveled output power is now available at the output of the directional coupler.
NOTE While operating in external leveling mode, the signal generator’s displayed RF output 
amplitude is affected by the coupling factor value, resulting in a calculated approximation of 
the actual RF output amplitude. To determine the actual RF output amplitude at the point of 
detection, measure the voltage at the external detector output and refer to Figure 4- 3 or 
measure the power directly with a power meter.
Determining the Leveled Output Power
Figure 4-3 shows the input power versus output voltage characteristics for typical Agilent 
Technologies diode detectors. Using this chart, you can determine the leveled power at the diode 
detector input by measuring the external detector output voltage. You must then add the coupling 
factor to determine the leveled output power. The range of power adjustment is approximately - 20 to 
+25 dBm.