92 Agilent X-Series Signal Generators User’s Guide
Optimizing Performance 
Using the Dual Power Meter Display 
13. On the signal generator: Press Channel B to On and then back to Off again. This initializes the 
signal generator to the external power meter.
14. Press  Return > Zero Sensor
A diagnostic dialog box is displayed each time an external power meter is being used and the 
Zero Sensor or Calibrate Sensor softkey is pressed (refer to Figure 5- 10 on page 92). 
Verify the power sensor is connected to the 50 MHz reference of the power meter.
Figure 5-10 Diagnostic Dialog Box for Channel B
15. Press  Done
The  Running Calibration(s) bar is displayed: Zeroing Sensor Please wait.... 
16. After Running Calibration(s) bar disappears: Press Calibrate Sensor 
Diagnostic Dialog box is displayed that prompts for verifying the connection of the power sensor 
to the power meter 50 MHz reference (refer to Figure 5- 11 on page 92).
Figure 5-11 Diagnostic Dialog Box for Calibration