94 Agilent X-Series Signal Generators User’s Guide
Optimizing Performance 
Using the Power Meter Servo 
Using the Power Meter Servo
The Power Meter Servo mode uses power meter readings to adjust the output power of the source, 
maintaining a constant DUT output power.
The servo loop measures the output power of the DUT, compares it to the user- provided reference 
power, and adjusts the output of the source to achieve the user- provided power level within the 
settling error. The servo loop will abort after twenty unsuccessful attempts to achieve the 
user- provided power level.
Figure 5-14 Power Meter Servo Menus
For details on each key, use key help as 
described on see page 44.
See page 88
Sets the allowed 
settling error.
The system automatically adjusts 
power level according to the power 
meter reading.
Once performs the adjustment only at 
the end of any transition (amplitude or 
frequency change). After the 
adjustment is performed and the 
power error is corrected, no further 
adjustments are performed until the 
next transition.
Continuous performs the adjustment 
as in Once mode, and continues to 
adjust the power periodically if the 
value differs by more than the 
specified Settling Error.
N5173B/83B only
N5173B/83B 
only
Sets the relative initial power 
when leveling with the Power 
Meter Servo mode. The 
default of 0 dB means that the 
system starts with the power 
it thinks is needed (based on 
output power, offset and 
reference) but this might lead 
to power overshoots in certain 
cases (especially when the 
offset is not set up correctly). 
This value allows to specify 
an additional safe-margin to 
start the measure/adjust 
cycles lower than the target 
power to protect the device 
from power overshoots.