108 Agilent X-Series Signal Generators User’s Guide
Optimizing Performance 
Using Internal Channel Correction (N5172B/82B Only) 
Figure 5-18 Internal Channel Correction Softkeys
SCPI Commands:
[:SOURce]:DM:INTernal:CHANnel:CORRection[:STATe] ON|OFF|1|0
[:SOURce]:DM:INTernal:CHANnel:CORRection[:STATe]?
:CALibration:BBG:CHANnel
:MEMory:LOAD:CHANnel <"filename"> 
:MEMory:STORe:CHANnel <"filename">
For details on each key, use key help as described on page 44.
I/Q > More
Toggles on or off the application of the 
internal baseband generator RF and 
baseband magnitude and phase 
corrections across the 160MHz 
baseband bandwidth at the current RF 
frequency.
When this feature is on, arbitrary 
frequency switching while the 
baseband is on will take up to an 
additional 3.3ms (typical) to 6.8ms the 
first time that frequency is specified. 
After that, switching to that frequency 
will take up to an additional 1.3ms. 
Up to 1024 unique frequencies can be 
cached before the oldest cache will be 
forgotten. If a frequency sweep is 
activated, then the calculation and 
caching will occur up front for the first 
1024 unique frequencies, and all 
further unique frequencies will have 
the characteristics of arbitrary 
frequency switching.
If the I/Q Correction Optimized Path is 
set to Ext I/Q Output then only the 
baseband corrections will be applied 
and the frequency switching speed will 
be unaffected.
This correction will be convolved with 
the ACP Internal I/Q Channel 
Optimization filter and the Equalization 
filter if they are active. The resulting 
filter will be truncated to 256 taps.
NOTE: The Enhanced Factory 
Calibration must be executed once 
before the Factory correction type can 
be turned on. See Perform Enhanced 
Factory Calibration.
Provides additional correction to the 
DUT.
Specify the start and stop frequencies, 
configure the power meter, then 
execute the calibration.
Displays a menu that controls the 
calibration and application of the 
internal baseband generator RF and 
baseband magnitude and phase 
corrections across the entire 
baseband bandwidth.
Toggles on or off the application 
of Factory or User I/Q channel 
correction. 
Factory optimizes the 
performance to the RF output 
connector.
User extends the correction to 
the DUT.
NOTE: The Enhanced Factory 
Calibration must be executed 
once before the Factory 
correction type can be turned on. 
See Perform Enhanced Factory 
Calibration.
To perform the Factory 
calibration, press Utility > More > 
Service > I/Q Int Channel Correction 
Calibrations.
Select either Factory Calibration or 
Enhanced Factory Calibration. 
Follow the instructions on the 
signal generator display.