User`s guide
118 Agilent X-Series Signal Generators User’s Guide
Optimizing Performance 
Using Unleveled Operating Modes 
Using Unleveled Operating Modes
Figure 5-22 Power Search and ALC Off Softkeys
Available only when
ALC = Off
Available only when
Power Search Reference =
Available only when 
Power Search = Span 
Auto: The calibration routine executes whenever 
output frequency or amplitude changes.
Span: Pressing Do Power Search executes the 
power search calibration routine once over a 
selected frequency range.
The corrections are stored and used whenever you 
tune the signal generator to within the calibrated 
frequency range.
For details on each key, use key help as described on see page 44.
RMS: Value is a DC 
bias equivalent to the 
value derived from the 
file header or the 
calculated value from 
the current I/Q data.
Fixed: Uses a fixed 0.5 
Vrms value (Same 
value is used with Ext 
I/Q data).
Manual: Allows user to 
specify the DC bias 
voltage value  
(0–1.414 Vrms).
Modulated: Uses the 
AC bias from the actual 
modulating signal.
Manual
User: Enables the 
user to set the 
signal generator 
frequency settings.
Full: The signal 
generator selects 
its full frequency 
range for the power 
search.
Only available when 
I/Q is on. 
These 
softkeys are 
only active 
when an ARB 
waveforms is 
playing in 
memory.
Manual: User must explicitly execute a power 
search to calibrate the open loop ALC level.  
See page 101
Sets the allowed 
settling error.
Only available when 
I/Q is on. 
N5173B/83B only
N5173B/83B only