User`s guide
 Agilent X-Series Signal Generators User’s Guide 135
 Using Pulse Modulation (Option UNW or 320)
Pulse Train (Option 320 – Requires: Option UNW)
Pulse Train (Option 320 – Requires: Option UNW) 
The Option 320 Pulse Train feature enables the specification of up to 2047 independent pulse cycles, 
each of which has an “On Time”, during which the RF output is measurable at the RF output 
connector, and an "Off Time", during which the RF output is attenuated. Each pulse cycle is similar 
in function to other X- Series signal generator Pulse modes—the Pulse Train feature has up to 2047 
cycles, instead of only a maximum of two (with Doublets). There are also repeat counts available for 
each pulse. These repetitions count against the total count of 2047 cycles.
The instrument can import pulse trains from a .csv (comma separated values) file or some other 
common ASCII format. It can also export to ASCII/CSV files as well. Export allows specification of 
the decimal separator and a column separator. The import allows specification of the decimal 
separator (to allow for ","), but the column separator is auto- detected. Refer to “Pulse Train Menu 
Softkeys” on page 135 and “Display Pulse Train Menu Softkeys” on page 137.
Figure 6-4 Pulse Train Menu Softkeys
Pulse > Pulse Source > More > Pulse Train 
Free-Run continuously plays 
the pulse train, ignoring all 
triggers.
Triggered runs the pulse 
train (after waiting the Pulse 
Delay) each time an external 
trigger is supplied to the 
PULSE BNC, the Trigger 
Immediately softkey is 
pressed, or the Trigger SCPI 
command is sent.
Gated runs the pulse train 
while an external trigger is 
supplied (level triggered) to 
the PULSE BNC connector. 
The state of the GATEd 
trigger is detected only when 
the playback is transitioning 
to or in idle. This means that, 
once started, playback is 
always completed, even if 
the GATE trigger changes to 
the inactive state.
For details on each key, use key help as described on page 44.
Display area indicates Pulse Train is the current pulse 
source.
Trigger Immediately causes 
the pulse train to run once.
SCPI Commands:
[:SOURce]:PULM:INTernal:TRAin:TRIGger FRUN|{TRIGgered}|GATEd
[:SOURce]:PULM:INTernal:TRAin:TRIGger:IMMediate  
Refer to the SCPI Command Reference.