User`s guide
142 Agilent X-Series Signal Generators User’s Guide
Basic Digital Operation—No BBG Option Installed
I/Q Modulation
I/Q Modulation
The following factors contribute to the error vector magnitude:
• Differences in amplitude, phase, and delay between the I and Q channels
•DC offsets
The I/Q menu provides adjustments and calibration to compensate for some of the differences in the 
I and Q signals or to add impairments. See I/Q Modulation on page 204 for additional information.
See also “Modulating the Carrier Signal” on page 59.
Figure 7-1 I/Q Display and Softkeys
This panel displays the external 
I/Q signal routing.
This panel displays the current status and settings 
of the I/Q adjustments. Grey text indicates I/Q 
adjustments are off.
For details on each key, use key help 
as described on page 44.
Offsets the phase of the Q signal 
relative to the phase of the I signal. 
The quadrature adjustment key is in 
units of degrees. This adjustment is 
not calibrated. 
Sets the dc 
offset
Toggles between DC, User, and Full IQ calibration types. 
DC: Calibrates I/Q offset, gain and quadrature error for a single 
frequency at the current signal generator settings. 
This calibration requires that the I/Q be on, the optimized path be set to 
RF Out, and the source be set to the internal baseband generator. 
Changing any instrument setting except for I/Q adjustments after 
performing a DC calibration voids the DC calibration and causes the 
signal generator to revert to the factory supplied calibration data. 
User: Calibrates I/Q offset, gain, and quadrature error for all signal 
generator settings over a range of frequencies that you define. 
Full: Calibrates I/Q offset, gain, and quadrature error for all settings 
over the full frequency range of the signal generator.
Starts the I/Q calibration 
routine for the selected 
frequency range. (Use the 
Calibration Type softkey to 
select the frequency range.) 
You can abort the 
calibration by pressing the 
Abort Cal softkey or the 
Local Cancel/(Esc) hardkey. 
If you abort the calibration, 
the previous calibration data 
is restored.
This calibration should be 
run when the ambient 
temperature has varied by 
at least +/-5 degrees 
Celsius from the ambient 
temperature at which the 
previous calibration was 
run.










