158 Agilent X-Series Signal Generators User’s Guide 
Basic Digital Operation (Option 653/655/656/657) 
Saving a Waveform’s Settings & Parameters 
The  Current Inst. Settings column shows the current signal generator settings. In this 
example, these are the settings that you will save to the file header. 
NOTE If a setting is unspecified in the file header, the signal generator uses its current value for 
that setting when you select and play the waveform.
Figure 8-5 Example File Header
3. Save the information in the Current Inst. Settings column to the file header:
Press  Save Setup To Header.
Both the Saved Header Settings column and the Current Inst. Settings column now display 
the same values; the Saved Header Settings column lists the settings saved in the file header.
4. Edit and Update Settings
a. Return to the ARB Setup menu:  
Press  Return > More > ARB Setup.
From this menu you can access some of the signal generator settings that are saved to the file 
header.  Figure 8- 1 on page 147 shows the ARB Setup softkeys used in the following steps.
b. Set the ARB sample clock to 5 MHz:  
Press  ARB Sample Clock > 5 > MHz.
c. Set waveform runtime scaling to 60%: 
Press  Waveform Runtime Scaling > 60 > %.
The name of the waveform file.
The description can be up to 
32–characters.
Mode > Dual ARB > More > Header Utilities
Default header settings
Current signal generator settings
Resets the saved 
header settings 
entries to default 
settings
To calculate the RMS waveform voltage: 
Press Calculate
Then refer to Header Field information.
Note: For greater accuracy, if the waveform has long rise/fall times, a DC 
offset, or noise* already added to the waveform (for a bursted signal), it 
is recommended that the Edit RMS and Edit AWGN RMS Override 
softkeys are used for the best measurement accuracy. 
*Option 403 Real–time AWGN does not affect the Header Field RMS 
value. But, noise added to the waveform prior to downloading and 
playing in the signal generator does affect the Header Field RMS value.
Opens a menu for manually 
defining the carrier RMS value 
to use for calculating the 
AWGN: Carrier RMS value in 
the Header Field. 