User`s guide
164 Agilent X-Series Signal Generators User’s Guide 
Basic Digital Operation (Option 653/655/656/657) 
Using Waveform Markers 
Close–up of averaging
The ALC samples the waveform when the marker signal goes 
high, and uses the average of the sampled waveform to set the 
ALC circuitry.
Here the ALC samples during the on marker points (positive 
polarity).
Marker
Marker
MarkerMarker
Example of Correct Use
Waveform: 1022 points 
Marker range: 95–97 
Marker polarity: Positive
This example shows a marker set to sample the waveform’s area of 
highest amplitude. Note that the marker is set well before the 
waveform’s area of lowest amplitude. This takes into account any 
response difference between the marker and the waveform signal.
Pulse Unleveled
Example of Incorrect Use
Waveform: 1022 points 
Marker range: 110–1022 
Marker polarity: Positive
This example shows a marker set to sample the low part of the 
same waveform, which sets the ALC modulator circuitry for 
that level; this usually results in an unleveled condition for the 
signal generator when it encounters the high amplitude of the 
pulse.
MarkerMarker