User`s guide
168 Agilent X-Series Signal Generators User’s Guide 
Basic Digital Operation (Option 653/655/656/657) 
Using Waveform Markers 
Press  Last Mkr Point > 17 > Enter > Apply To Waveform > Return.
This turns off all marker points for 
the active marker within the range 
set in Steps 2 and 3, as shown at 
right.
How to view markers is described 
on  page 167.
Clearing a Single Marker Point
Use the steps described in “Clearing a Range of Marker Points” on page 167, but set both the first 
and last marker point to the value of the point you want to clear. For example, if you want to clear 
a marker on point 5, set both the first and last value to 5.
Setting Marker Points in a Waveform Segment
To set marker points on a segment, the segment must reside in BBG media (page 148).
When you set marker points, they do not replace points that already exist, but are set in addition to 
existing points. Because markers are cumulative, before you set marker points within a segment, view 
the segment (page 167) and remove any unwanted points (page 167).
Placing a Marker Across a Range of Points
1. In the second Arb menu (page 166), press Marker Utilities > Set Markers.
2. Highlight the desired waveform segment.
3. Select the desired marker number: Press Marker 1 2 3 4
4. Set the first sample point in the range (in this example, 10):
Press  Set Marker On Range Of Points > First Mkr Point > 10 > Enter.
5. Set the last marker point in the range to a value less than or equal to the number of points in 
the waveform, and greater than or equal to the first marker point (in this example, 20):
Press  Last Mkr Point > 20 > Enter.
6. Press  Apply To Waveform > Return.
This sets a range of waveform marker points. The marker signal starts on sample point 10, and ends 
on sample point 20, as shown in the following figure.
How to view markers is described on page 167










