User`s guide
 Agilent X-Series Signal Generators User’s Guide  177
 Basic Digital Operation (Option 653/655/656/657)
Using Waveform Markers
Using the EVENT Output Signal as an Instrument Trigger
One of the uses for the EVENT 
output signal (marker signal) is to 
trigger a measurement instrument. 
You can set up the markers to start 
the measurement at the beginning 
of the waveform, at any single point 
in the waveform, or on multiple 
points in the waveform. To optimize 
the use of the EVENT signal for 
measurements, you may also need 
to adjust the sample rate. The 
location of the sample rate setting 
is shown in the figure at right.
The EVENT output signal can exhibit jitter of up to ±4 ns on the rising and falling edge. This jitter 
can be minimized in either of two ways. 
Method 1: Use a sample clock of 125 MHz/N where N is a positive integer and where 125 MHz/N can 
be represented exactly on the display.  
For example: 125 MHz, 62.5 MHz, 31.25 MHz, 25 MHz, and so on. 
If the result cannot be represented exactly on the display, jitter will be present. 
For example: N = 6 will result in jitter, because 125 MHz/6 = 20.833 Mhz, which is truncated when 
displayed.
Method 2: Select a sample clock and waveform length that spaces the markers by a multiple of 8 ns. 
For example: A 200 point waveform with a marker on the first point and a sample clock of 50 MHz 
provides a marker every 4 µs. Because 4 µs is a multiple of 8 ns, the jitter is minimized.
When the EVENT output signal exhibits jitter and it is used as a measurement trigger, it can cause 
the waveform to falsely appear as having jitter. If this condition occurs, you can adjust the sample 
rate to a value (see above) that does not cause the jitter appearance. To maintain the integrity of the 
original waveform with a sample rate change, you will have to also recalculate the sample values. The 
following figures illustrate the marker signal jitter and its affect on the waveform.
Sample 
rate  
setting
The settings in this menu 
can be stored to the file 
header, see page 155.
For details on each key, use key help as described on page 44.
The jitter is gone with
an optimal sample rate
Waveform appears to exhibit jitter when
triggered using EVENT signal with jitter.
Oscilloscope triggering on waveform
Oscilloscope triggering on EVENT signal
Oscilloscope triggering on EVENT signal
EVENT output signal exhibits jitter
due to a non–optimal sample rate