 Basic Digital Operation (Option 653/655/656/657)
Setting the Baseband Frequency Offset
Figure 8-14 Dual ARB DAC Over–Range Protection Softkey Location
In the Dual ARB Player, to avoid excessive scaling or to just perform scaling manually, turn the 
feature off and use the Waveform Runtime Scaling softkey to eliminate DAC over–range conditions.
For details on each key, use key help 
as described on page 44.
Default setting is On. Available only 
when Phase Noise or AWGN are on 
(see page 309), or when the 
baseband frequency offset is set to a 
value other than zero.
Turn off when you want to manually 
control scaling while using the 
baseband frequency offset feature. 
Manually adjust the scaling using the 
Waveform Runtime Scaling softkey 
above.
Enables or disables the automatic 
internal scaling used to keep DAC 
over-range errors from occuring with 
the Phase Noise, AWGN and 
Frequency Offset features. This 
automatic scaling can be excessive 
(particularly for a constant amplitude 
signal) and thus reduce the dynamic 
range more than is necessary.
When the protection is disabled, no 
automatic scaling occurs, so the 
Runtime Scaling feature must be 
used to adjust for the highest scaling 
value that does not cause DAC 
overranges.
Since the over-range protection 
works with only three features, this 
key grays out until one of the three 
features is active even when the 
protection is on. 
When the DAC over–range protection is off, 
eliminate over–range conditions by 
decreasing the scaling value (see “Setting 
Waveform Runtime Scaling” on page 197).
page 226