User`s guide
 Agilent X-Series Signal Generators User’s Guide  271
 Digital Signal Interface Module (Option 003/004)
Operating the N5102A Module in Output Mode
Choosing the Logic Type and Port Configuration
Figure 10-11Logic and Port Configuration Softkey Menus
1. Refer to Figure 10- 11. Press the Logic Type softkey.
From this menu, choose a logic type.
CAUTION Changing the logic type can increase or decrease the signal voltage level going to the 
device under test. To avoid damaging the device and/or the N5102A module, ensure 
that both are capable of handling the voltage change.
2. Select the logic type required for the device being tested.
A caution message is displayed whenever a change is made to the logic types, and a softkey 
selection appears requesting confirmation.
3. Refer to Figure 10- 11. Press the Port Config softkey.
In this menu, select either a serial, parallel, or parallel interleaved data transmission.
NOTE Within the data and clock setup softkey menus, some softkeys function relative to the 
current configuration. Softkeys that are grayed out are not available for the current 