User`s guide
 Agilent X-Series Signal Generators User’s Guide  289
 Digital Signal Interface Module (Option 003/004)
Operating the N5102A Module in Input Mode
2. Press the Data Type softkey. 
In this menu, select the data type to be either filtered (Samples) or unfiltered (Pre-FIR Samples). 
The selection is dependent on the test needs and the device under test. However if the device 
being tested already incorporates FIR filters, the Pre-FIR Samples selection should be used to avoid 
double filtering. Refer to 
“Data Types” on page  268, for more information.
3. Select the data type that is appropriate for the test needs.
4. Press the Numeric Format softkey.
From this menu, select how the binary values are represented. Selecting 2’s complement allows 
both positive and negative data values. Use the Offset Binary selection when components cannot 
process negative values.
5. Select the numeric format required for the test.










